Popular application areas:
6N hydrogen (H2), argon (Ar), nitrogen (N2), helium (He), carbon dioxide (CO2), various gas mixtures
Electronic Grade Gas Analyzers
Product Features :
I、Multi Mix Gas Analyzer Functional Specifications
- Types of gases that can be analyzed:hydrogen (H2), argon (Ar), helium (He), nitrogen (N2), carbon dioxide (CO2), various mixed gases.
- The composition of the analyte and the lowest detectable concentration (QDL):H2, Ar, N2, CH4, CO, CO2, N2O. QDL is <10ppb.
- Total Hydrocarbons (THC) Minimum Detectable Concentration (QDL):<10ppb.
- The relative standard deviation (RSD) of the five-repeated analysis of the standard composition of the analyte (5ppm):1~2%.
II、Multi Mix Gas Analyzer Hardware Specifications
A. Chromatography System
- Electronic Flow Control (CGM):Standard specification for carrier gas, detector flow, auxiliary gas, backflush gas, and heart cut gas.
- Oven temperature operating range and accuracy:35℃~200℃.
- It can be configured with 3 independent temperature-controlled incubators, 3 detectors and 3 sets of independent temperature-controlled sample injection systems.
- Finger-Tight style string connection, no tools required for string replacement.
- Injection volume:0.06ul~5ml.
B. Detectors and Control Systems
1. Pulsed Discharge Detector (PDHID):
- Universal non-destructive helium photoionization detection mode.
- Helium photon energy:13.5~17.7eV.
- 最Minimum detection volume:low ppb and can analyze non-corrosive or corrosive gases.
- Linear range:greater than 105
2. Flame Ionization Detector (FID):
- Electronic Flow Control (EPC):Controls the flow of hydrogen, air, make-up gas.
- Minimum detection:less than or equal to 1.8pg (10-12g) carbon/sec. Linear dynamic range:≧107
3. Thermal Conductivity Detector (TCD):
- Electronic Flow Control (EPC):Controls the flow of reference gas, make-up gas.
- Minimum detection volume:better than 400x10-12g hydrocarbons/ml. Linear dynamic range 105.
- Programmable setting of signal change output time.
C. Output Signal
- Analog output 0-1V or 4-20mA(option).
- Various communication protocols.