High Purity Oxygen (O2) Gas Analyzer

Product Features :

Popular application areas:
Steel, Air Separation Plant, High Purity Gas, High Purity Oxygen, O2


  Oxygen (O2) Trace Component Gas Analyzer 

 
Oxygen supply plays an important role in the steelmaking process, semiconductor process and liquid crystal display process. The purity of oxygen dominates the yield of each process product and the risk of industrial safety. In the 1990s, air separation equipment explosion casualties occurred in foreign countries, and the safety of air separation equipment has attracted great attention. The current online detection of hazardous gas components in the liquid oxygen process in the oxygen factory relies on old equipment with limited functions such as GC-FID and NDIR, and the analysis capability is insufficient. Therefore, this hazardous component detection instrument was developed to analyze the risks of the liquid oxygen process in the oxygen factory online. Explosive gas components CO2, N2O, CH4, C2~C4, impurity components Ar, Kr, N2, trace analysis technology can reach below 50ppb, improve and expand the applicability of the analyzer to achieve the following expected goals.
 
    ▶  The new analytical detection technology that measures all components at once replaces the traditional three analytical detection techniques.
    ▶  Accurately detect trace and micro-explosive components to improve the industrial safety of oxygen fields.
 

  I、Oxygen (O2) Gas Analyzer Functional Specifications   

 

  • Analyzable gas species:Oxygen.
  • The composition of the analyte and the minimum detectable concentration (MDL):Ar, N2, Kr, CH4, CO2, N2O, C2~C4.  MDL is 30~100ppb.
  • The relative standard deviation (RSD) of the standard composition of the analyte (5ppm) in five repeated analysis:<3~6%.

 

  II、Oxygen (O2) Gas Analyzer Hardware Specifications  

 

   A. Chromatography System

 

  • Electronic Flow Control (EPC):Accurate to 0.001psi, it is the standard specification for carrier gas, detector flow, auxiliary gas, backflush gas and core cut gas (heart cut). It can control 16 channels of flow flow or pressure.
  • Analytical component retention time error:<0.0008 minutes.
  • Oven temperature operating range and accuracy:room temperature +4 ℃ ~ 450 ℃, 0.01 ℃ control accuracy.
  • 3 detectors can be configured.

 

   B. Detectors and Control Systems

 

     1. Pulse Discharge Detector(PDHID):

  • Universal non-destructive helium photoionization detection mode.
  • Helium photon energy:13.5~17.7eV.
  • Minimum detection volume:low ppb and can analyze non-corrosive or corrosive gases.
  • Linear range:greater than 105.

 

     2. Flame Ionization Detector(FID):

  • Electronic Flow Control (EPC):Controls the flow of hydrogen, air, make-up gas.
  • Minimum detection:less than or equal to 1.8pg (10-12g) carbon/sec, linear dynamic range:≧107.
  • Total Hydrocarbons (THC):Minimum detection < 0.01 ppm.
 

    3.  Thermal Conductivity Detector(TCD):

  • Electronic Flow Control (EPC):Controls the flow of reference gas, make-up gas.
  • Minimum detection volume:better than 400x10-12g hydrocarbons/ml, linear dynamic range 105.